SPMK 51X Series Process Calibrator

SPMK 51X Series Process Calibrator

Short Description:

SPMK 51X Process Calibrator:

1. SPMK518 Intelligient Process Calibrator (HART / PRESSURE / TEMPERATURE)

2.SPMK517 Process Calibrator (PRESSURE / TEMPERATURE)

3.SPMK516 Process Calibrator (TEMPERATURE)

4. SPMK510 Loop Calibrator

 


Operation Video

Product Detail

OVERVIEW:

The SPMK51x series calibrator is with 3.5 Inch TFT color screen, cellphone-like menu, HART communication capability and powerful task management function. It is ideal for calibrating, maintaining, and troubleshooting process instrumentation.

Features:

Sources and measures mV, mA, ohms, frequency, pulses and switch, also can do temperature-test with thermocouples and RTDs independently

Simulates and measures 13 thermocouples (K E J T N R S B C D G L U) and 11 RTDs (Pt1000(385),Pt500(385),Pt100(385),Pt100(3916),Pt100(3926),Pt10(385)

Cu10(427),Cu50(428),Cu100(428),Ni120(672),Ni100(618)) with automated cold junction compensation

24V loop power supply with independent measuring and sourcing function

Documenting storage and automated procedure capability

Connect with SPMK SDP intelligent pressure modules

HART communication capability

With RS232 interface

      Product SPMK518 SPMK517 SPMK516 SPMK510
 
Function Intelligent process calibrator Multifunction process calibrator Multifunction temp calibrator Multifunction loop calibrator
Measurement Millivoltvoltage
Voltage
Current
Frequency
Pulse
Resistance
On-off
11 RTDs
13 thermocouples
Output Millivolt voltage
Voltage
Current
Frequency
Pulse
Resistance
Simulate11 RTDs
Simulate13 thermocouples
Other Functions HARTcommunication
Externalpressuremodule ①
Thermal book
Step &programmingoutput
24V loop power supply
Simulating transmitter
RS232 communication
Storage
Signal Range Accuracy Resolution
Current (-30-30)mA ±(0.01%RD+0.005%F.S) 0.1uA
Voltage (-30-30)V ±(0.01%RD+0.005%F.S) 0.1mV
mV (-75-75)mV ±(0.01%RD+0.005%F.S) 0.1uV
R-3W (0-400)Ω ±(0.02%RD+0.005%F.S) 1mΩ
(0-4000)Ω
R-4W (0-400)Ω ±(0.01%RD+0.005%F.S) 1mΩ
(0-4000)Ω
Frequency (1-50)KHz ±(0.005%RD+0.002%F.S) 0.1Hz
Pulse 0-999999 ±1pcs 1
Limit Switch If the switch has detective voltage, its range is (3-24)V
Thermocouple K E J T N R S B C D G L U
RTD Pt1000(385),Pt500(385),Pt100(385),Pt100(3916),Pt100(3926),Pt10(385)
Cu10(427),Cu50(428),Cu100(428),Ni120(672),Ni100(618)
Signal Range Accuracy Resolution
Current (0-22)mA ±(0.02%RD+0.005%F.S) 0.1uA
Voltage (0-12)V ±(0.02%RD+0.005%F.S) 0.1mV
mV (-10-100)mV ±(0.02%RD+0.005%F.S) 1uV
Resistance (0-400)Ω ±(0.02%RD+0.005%F.S) 1mΩ
(400-4000)Ω ±(0.03%RD+0.01%F.S) 10mΩ
Frequency (0-50000)Hz ±(0.005%RD+0.002%F.S) 0.1Hz
Pulse 0-999999 ±1pcs 1
DC24V 24.0±0.5V N/A N/A
Thermocouple K E J T N R S B C D G L U
RTD Pt1000(385),Pt500(385),Pt100(385),Pt100(3916),Pt100(3926),Pt10(385)
Cu10(427),Cu50(428),Cu100(428),Ni120(672),Ni100(618)

  • Previous:
  • Next:

  • Related Products